Forced air warming system: evaluation of internal system contamination (2019)
- Authors:
- Autor USP: POVEDA, VANESSA DE BRITO - EE
- Unidade: EE
- Sigla do Departamento: ENC
- DOI: 10.1089/sur.2018.152
- Subjects: HIPOTERMIA; INFECÇÃO HOSPITALAR; CIRURGIA; ENFERMAGEM PERIOPERATÓRIA
- Language: Inglês
- Abstract: Background: Despite the indisputable need to prevent peri-operative hypothermia, some evidence in the literature questions the possible role of forced air warming systems as a risk factor for the occurrence of surgical site infection. The objective of the study is to evaluate the micro-biologic safety of a forced air warming systems (FAW) in relation to the risk of emission of micro-organisms in the surgical environment. Methods: A quantitative, descriptive-exploratory laboratory study performed in a large hospital. An evaluation of possible internal contamination was performed by collecting air from the hose onto plates containing Trypticase Soy Agar. The experiment was performed in triplicate two months after replacing the filter and again after six months using a random sample of 50% of the FAW in use, which corresponded to 13 FAW and a total of 75 samples. Results: Among the 39 samples analyzed in the first stage, only nine (23.1%) plates presented microbial growth of one or two colonies, while only six (16.7%) plates of the 36 evaluated samples in the second phase had growth of one or two colonies. Conclusions: This study showed small microbial growth of culture after 48 hours after filter replacement. New investigations that correlate the findings of micro-biology analysis and the occurrence of surgical site infection should be conducted
- Imprenta:
- Source:
- Título do periódico: Surgical infections
- ISSN: 1096-2964
- Volume/Número/Paginação/Ano: v. 20, p. 1-4, 2019
- Este periódico é de assinatura
- Este artigo NÃO é de acesso aberto
- Cor do Acesso Aberto: closed
-
ABNT
SILVA, Letícia Lopes da et al. Forced air warming system: evaluation of internal system contamination. Surgical infections, v. 20, p. 1-4, 2019Tradução . . Disponível em: https://doi.org/10.1089/sur.2018.152. Acesso em: 19 abr. 2024. -
APA
Silva, L. L. da, Almeida, A. G. C. dos S., Possari, J. F., & Poveda, V. de B. (2019). Forced air warming system: evaluation of internal system contamination. Surgical infections, 20, 1-4. doi:10.1089/sur.2018.152 -
NLM
Silva LL da, Almeida AGC dos S, Possari JF, Poveda V de B. Forced air warming system: evaluation of internal system contamination [Internet]. Surgical infections. 2019 ; 20 1-4.[citado 2024 abr. 19 ] Available from: https://doi.org/10.1089/sur.2018.152 -
Vancouver
Silva LL da, Almeida AGC dos S, Possari JF, Poveda V de B. Forced air warming system: evaluation of internal system contamination [Internet]. Surgical infections. 2019 ; 20 1-4.[citado 2024 abr. 19 ] Available from: https://doi.org/10.1089/sur.2018.152 - Vigilância pós-alta em infecção de sítio cirúrgico: validação de um instrumento
- Hipotermia no período intra-operatório
- Cultura de segurança do paciente: percepção de residentes e apromorandos de um hospital cardilógico
- Diagnósticos de enfermería en pacientes sometidos a hemodiálisis
- Contaminação microbiana de punhos de jalecos durante a assistência à saúde
- Validation of an instrument for postoperative survellance in surgical site infection
- Effects of intensive blood glucose control on surgical site infection for liver transplant recipients: a randomized controlled trial
- Effectiveness of combined virtual and clinical simulation compared with other active teaching strategies on health students’ learning: a systematic review protocol
- vigilância pós-alta: uma revisão integrativa
- Mediastinite no pós-operatório de cirurgia cardíaca: estudo transversal retrospectivo
Informações sobre o DOI: 10.1089/sur.2018.152 (Fonte: oaDOI API)
Download do texto completo
Tipo | Nome | Link | |
---|---|---|---|
POVEDA, V de B doc 38.pdf | Direct link |
How to cite
A citação é gerada automaticamente e pode não estar totalmente de acordo com as normas